Quality and regulation
Built as a medical device, from the first commit.
Allelyn supports clinical decisions, so it is treated as Software as a Medical Device and developed inside a quality management system from day one.
Frameworks
ISO 13485Design controls, requirement traceability and change management.
IEC 62304Software life cycle. Safety class B is expected, to be confirmed.
ISO 14971Risk management: wrong sample, missed variant, wrong class, wrong report.
IEC 62366-1Usability engineering, especially on interpretation and sign-off screens.
EU IVDR (2017/746)The reference framework. Turkish regulation is aligned with it, and CE marking eases registration in MENA and Africa.
TİTCKFirst market: registration in Türkiye and notification in ÜTS.
ISO 15189Features that support customer laboratories' accreditation.
ISO 27001Information security management.
Clinical validation
Every pipeline version is validated against GIAB reference samples (NA12878, HG002–HG007), the laboratory's known positives and synthetic variants. Target for SNVs: sensitivity and positive predictive value ≥ 99.5%. Thresholds for indels and CNVs are set in the validation plan.
Security and privacy
- Under KVKK, genetic data is special-category personal data: explicit consent, data inventory (VERBİS) and cross-border transfer limits are designed in.
- Encryption at rest and in transit, with keys under the laboratory's control (KMS/HSM).
- Least privilege, field-level masking and pseudonymisation for research exports.
- Immutable, signed audit trail; electronic signatures designed to 21 CFR Part 11 and eIDAS principles.
- Annual penetration tests, SBOM and dependency scanning, disaster recovery with RPO ≤ 1 hour and RTO ≤ 8 hours.
Risks we design against
| Risk | Impact | Mitigation |
|---|---|---|
| Sample mix-up | Report on the wrong patient | Barcode chain, SNP fingerprint, trio parentage check |
| Missed pathogenic variant | False-negative diagnosis | Validation, low-coverage region report, limitations text |
| Wrong classification | Wrong clinical decision | Four-eyes approval, in-house knowledge base, reclassification flow |
| Interpretation that doesn't scale | Long turnaround | Phenotype prioritisation, automatic ACMG proposals |